Checking The Radiation Time - Sirona Orthophos XG 3D/Ceph Installation Manual

Table of Contents

Advertisement

12 Startup, measurements and tests (for USA/Canada only)
12.6 Checking the radiation time
Preparing the measurement
A
A
-
60kV/8mA
-
0.5s
Performing measurements
Analyzing measurements
Finalizing the measurement
184
12.6

Checking the radiation time

1. Attach the Mult-O-Meter sensor in the middle of the combisensor (on
2D sensor side).
2. Set the main switch (A) to I (see also Operating Instructions).
3. Wait for approx. 1 minute.
4. Press the R key.
 The unit moves to its starting position.
5. Call the Service menu and the Service routine S002.3 (see Service
A
Manual).
6. Use the arrow keys (A) in
60 kV/8 mA.
7. Use the arrow keys (A) in
0.5 s.
+
S2
+
3
➢ Initiate the radiation. Hold the release button pressed until the set
radiation time has expired.
➢ Read the radiation time on the Mult-O-Meter.
 The value for the radiation time displayed on the Mult-O-Meter
must correspond to the radiation time of 0.5s selected in the
service routine. The permissible tolerance is ± 10 %.
 If the measured radiation time does
tolerance, replace the
 If the measured radiation time falls within the permissible
tolerance, finalize the measurement.
1. Exit the service routine.
2. Switch the unit on via switch (A) (see also Operating Instructions).
Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS XG 3D / Ceph
selection field 1
to select the kV/mA level
selection field 2
to select the radiation time
not
fall within the permissible
tube assembly
(see Service Manual).
D3352.031.05.08.02 10.2015