S001: Radiation Without Rotary Movement, Fixed Maximum Radiation Time; Maximum Radiation Time - Sirona Orthophos XG 3D Ceph Service Manual

Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
SR*
S001
63 03 510 D3352
D3352.076.03.11.02 09.2016
8.4
S001: Radiation without rotary movement, fixed

maximum radiation time

SHZ**
Function
Yes
X-ray beam test
kV/ma levels test
Explanation:
● Radiation with selected kV/mA level
● Radiation time cannot be selected, max. 14.0 s
● Primary diaphragm remains in last position selected
* SR=service routine, ** SHZ=security access
Unit is radiating X-rays
Excess exposure to X-rays is detrimental to health.
➢ Use the prescribed accessories for radiation protection.
➢ Do not stay in the X-ray room during exposure. Move as far away
from the unit as the coiled cable for the release button allows you to.
Selection field
1
2
* Factory setting
1. Call service routine S001.
The maximum radiation time is displayed in selection field 2. The
maximum radiation time cannot be changed in this service routine. It
is set to 14.0 seconds. Radiation is emitted as long as the release
button remains pressed, but for no longer than 14.0 seconds.
2. Use the arrow keys (A) in selection field 1 to select the required kV/
mA level (see table).
8 Perform service routines via the control panel
8.4 S001: Radiation without rotary movement, fixed maximum radiation time
WARNING
Parameter/Display
kV/mA level
Radiation time
Range of values
60 kV/8 mA* –
85kV / 7mA
14s
263