5 Operation
5.4 Panoramic exposure
P1
14.1 64 8
Prog.
Prog.
S
T
R
A
P1
14.1 64 8
Prog.
Prog.
S
T
R
38
5.4.4

Setting the kV/mA values

Setting the kV/mA values via the patient symbols
Preset kV/mA combinations are assigned to the patient symbols, which
are selected according to the patient's size and weight. The symbols
roughly correspond to child, youth/woman, woman/man, hefty persons.
➢ Select the desired patient symbol. Press one of the four patient
symbol keys (A).
kV
mA
The kV/mA value is set.
Manual setting of kV/mA values
If the preset kV/mA combinations do not provide a satisfactory result, you
can also set the kV/mA values manually in all programs.
➢ Select another kV/mA value. Push the kV/mA keys forward (B) and
B
backward (C).
The kV/mA value is set.
kV
mA
C
The LED above the selected patient symbol lights up. The
corresponding kV/mA values appear on the digital display.
The selected kV/mA value is shown on the digital display. If the
new value happens to agree with the value programmed for
another patient symbol key, its LED then lights up.