4 Maintenance by the service engineer
4.9
Diaphragm check
Perform diaphragm check also with Cephalometer, if present.
Performing diaphragm check
Maintenance interval
(after years)
Date
(enter month/year)
Test phantom exposure/needle
phantom exposure correct?
See Service Manual
4.10
Checking the cables for damage
Checking the cables
Maintenance interval
(after years)
Date
(enter month/year)
Power cable OK?
See Service Manual
Protective ground wire OK?
Control cable OK?
Data transfer cable OK?
24
Answer questions with yes () or no (−)
4
7
10
If no Correct fault.
Answer questions with yes () or no (−)
4
7
10
If no Correct fault.
12
14
12
14