S017.17: Checking and accepting the diaphragm configuration, worldwide
Tab 5
Selecting test step 17
Case A:
The data of boards DX11 and DX61 agree.
Case B:
The data of boards DX11 and DX61 do not
match
Data must be transferred from
DX61 to DX11.
59 38 399 D3352
D3352.076.01.26.02 09.2016
5.13.15

S017: Test step 17

Checking and accepting the diaphragm con-
figuration, worldwide
(not for XG 3 / 3 PPE)
1.
Use the arrow keys in selection field 2 to select test step 17 and confirm
your selection via the Service key
Once this service routine has been selected, the data of boards DX61 and
DX11 are checked for agreement.
Plus
ready
XG
/ 3D
:
Easypad touchscreen
1
01
2
3
No further action is required.
Plus
ready
XG
/ 3D
:
Easypad touchscreen
1
01
61
2
11
3
T
R
2.
To transfer data from board DX61 to board DX11 and save them there,
first press the memory key
R key
and then the
R
5.13 Service routine S017
.
XG 5: Multipad
S17
______01________
17
1
XG 5: Multipad
S17
01____61 11_____
17
1
(R key or LED above R key lights up)
.
R
2
3
2
3
5 – 115