3.7 Using the menu system

The DAC UNIVERSAL has a menu system, which gives the user different options. The different options and how to use the
menu system will be explained below.
To enter menu level 1, press the menu button, press Up or Down button until the desired option is selected, and press
Enter to select.
At each menu level, press Up/Down until the desired option is selected, and press Enter to select.
Menu level 1
Menu level 2
Program
134C
121C
134C, 18 min
Manual
Open
Close
Water
Drain tank
Drain chamber
Settings
Printer
Rinse cycles
Oil expel
Language
Expel
Set time/date
Automatic water
Oil settings
60 53 362 D3472
D3472.201.01.08.02
DAC UNIVERSAL – USER MANUAL
729050 - Version
Menu level 3
On
Off
Rinse cycles (2-6)
Oil expel (1-5)
UK
D
F
SP
I
S
Expel (0-3)
Set year (00-99)
Set month (01-12)
Set day (01-31)
Time scale (12h/24h)
Set hour (01-24)
Set min (00-59)
On
Off
Oil settings (1-3)
16
S
Result/action
Sterilization program 134 Celsius for 3 minutes
Sterilization program 121 Celsius for 15 minutes
Sterilization program 134 Celsius for 18 minutes
Opens chamber lid
Closes chamber lid (remember to place the lid into the
holder (semi-circle) of the autoclave)
Pumps water into the chamber
Drains the water tank out through the water inlet.
Drains the chamber into the drain outlet.
Remember to place the lid into the holder (semi-circle)
before the use of this function.
App. 60 sec. after having chosen Drain Chamber and
pressed Menu/Enter - the lid goes up, and the chamber
should be empty. The function can be repeated, if the
chamber is not empty, after previous attempts.
Turns on printer reporting
Turns off printer reporting
Default is set at 2 pulse washes/ rinse cycles, but it is
possible to increase it up to 6, if this is required.
(please note that more than two Rinse cycles are
prolonging total time of cycle)
Default is set at level 1, but it is possible to choose up
to level 5, if a longer time for blowing air through the
instruments is required.
(please note that this is before sterilization and is
prolonging total time of cycle)
English is selected as display language
German is selected as display language
French is selected as display language
Spanish is selected as display language
Italian is selected as display language
Swedish is selected as display language
Default is set at level 0 – (NO water expel), but it is
possible to choose up to level 3, if a longer time for
blowing air through the instruments is required.
Level 0: Expel off
Level 1: 50 sec. effective blowing time
Level 2: 100 sec. effective blowing time
Level 3: 150 sec. effective blowing time
(please note that this is after sterilization and is
prolonging total time of cycle)
Sets year
Sets month
Sets day
Sets time format
Sets hour
Sets minute
Turns water on
Turns water off
Default is set at level 1 – (oil adjustment), but it is
possible to choose up to level 3, if a larger amount of
oil/lubrication for the instruments is requested.
Level 1: default amount
Level 2: 25% more than level 1
Level 3: 50% more than level 1
32