Operation Mode (Cvu) - Keithley 4200A-SCS Reference Manual

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Model 4200A-SCS Parameter Analyzer Reference Manual

Operation Mode (CVU)

The following topics describe the operation modes that are available when a CVU is selected as the
instrument.
Voltage Bias operation mode - CVU
The Voltage Bias operation mode maintains a selected constant-voltage state at the terminal.
The parameters that are available for this mode are briefly described in the following table. Select the
links to access additional information. The parameters are listed in the order in which they appear in
the All Parameters pane.
Parameter
Presoak
(on page 6-55)
DC Bias
(on page 6-57)
Frequency
AC Drive Signal
Parameters
Param1 Column Name
58)
Param2 Column Name
58)
Report Test Conditions
58)
DCV Column Name
Freq Column Name
Report Status
Compensation Open
Compensation Short
Compensation Load
Compensation Cable Length
page 6-60)
AC Source V
4200A-901-01 Rev. C / February 2017
(on page 6-57)
(on page 6-57)
(on page 6-58)
(on page 6-
(on page 6-
(on page 6-
(on page 6-59)
(on page 6-59)
(on page 6-59)
(on page 6-59)
(on page 6-59)
(on page 6-60)
(on
(on page 6-60)
Description
Type a presoak voltage of −30 V to +30 V.
The DC bias range accepts a value between −30 V to 30 V.
Select the frequency from the list of values.
The AC drive signal accepts a value between 10 mV and 100 mV.
Select the type of parameters. When using any of the tests supplied by
Keithley Instruments, leave the measurement option set to Cp-Gp.
The name that is used for parameter 1 in the Analyze sheet.
The name that is used for parameter 2 in the Analyze sheet.
Select this option to display the DC bias and drive frequency values on the
Analyze sheet.
Available if you select Report Test Conditions. Determines the name of the
column that contains the DC bias information in the Analyze sheet where
this data is reported. You can change the name of the column by typing a
new value.
Available if you select Report Test Conditions. Determines the name of the
column that contains the frequency information in the Analyze sheet where
this data is reported. You can change the name of the column by typing a
new value.
When this option is selected, Clarius records measurement status
information when the test executes. A column of the Analyze spreadsheet
displays this information. Hover over a cell to review the information.
Use the CVU connection compensation value that was generated for open
connection compensation. Refer to
10) for information on generating and using compensation values.
Use the CVU connection compensation value that was generated for short
connection compensation. Refer to
10) for information on generating and using compensation values.
Use the CVU connection compensation value that was generated for load
compensation. Refer to
Connection compensation
information on generating and using compensation values.
The cable length that was used to generate connection compensation data.
Make sure this cable length is the same as the Cable Length setting in the
Tools > CVU Connection Compensation dialog box. Refer to
compensation
(on page 4-10) for additional information.
Selects the terminal to use to source AC drive voltage.
Section 6: Clarius
Connection compensation
(on page 4-
Connection compensation
(on page 4-
(on page 4-10) for
Connection
6-47

Advertisement

Table of Contents
loading

Table of Contents