Keithley 4200A-SCS Reference Manual page 323

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Section 6: Clarius
The parameters that are available for this mode are briefly described in the following table. Select the
links to access additional information. The parameters are listed in the order in which they appear in
the All Parameters pane.
Parameter
Presoak
(on page 6-55)
Start
(on page 6-55)
Stop
(on page 6-55)
Step
(on page 6-36)
Points
(on page 6-38)
Start Frequency
Stop Frequency
Frequency Points
AC Drive Signal
Parameters
(on page 6-58)
Param1 Column Name
58)
Param2 Column Name
58)
Report Test Conditions
58)
DCV Column Name
Freq Column Name
Report Status
Compensation Open
Compensation Short
Compensation Load
6-54
Description
Type a presoak voltage of −30 V to +30 V.
The voltage source level at which the step starts.
The voltage source level at which the step stops.
The voltage size of each step of the sweep. The source level changes in
equal steps of this size from the start level to the stop level. A measurement
is made at each source step (including the start and stop levels).
The number of data points that will be measured. This value is calculated by
Clarius using the information entered for the Start, Stop, and Step
parameters.
(on page 6-56)
The frequency at which the sweep starts.
(on page 6-56)
The frequency at which the sweep stops.
(on page 6-56)
The number of frequency points to sweep. This is calculated from the start
and stop frequency values.
(on page 6-57)
The AC drive signal accepts a value between 10 mV and 100 mV.
Select the type of parameters. When using any of the tests supplied by
Keithley Instruments, leave the measurement option set to Cp-Gp.
(on page 6-
The name that is used for parameter 1 in the Analyze sheet.
The name that is used for parameter 2 in the Analyze sheet.
(on page 6-
(on page 6-
Select this option to display the DC bias and drive frequency values on the
Analyze sheet.
(on page 6-59)
Available if you select Report Test Conditions. Determines the name of the
column that contains the DC bias information in the Analyze sheet where
this data is reported. You can change the name of the column by typing a
new value.
(on page 6-59)
Available if you select Report Test Conditions. Determines the name of the
column that contains the frequency information in the Analyze sheet where
this data is reported. You can change the name of the column by typing a
new value.
(on page 6-59)
When this option is selected, Clarius records measurement status
information when the test executes. A column of the Analyze spreadsheet
displays this information. Hover over a cell to review the information.
(on page 6-59)
Use the CVU connection compensation value that was generated for open
connection compensation. Refer to
10) for information on generating and using compensation values.
(on page 6-59)
Use the CVU connection compensation value that was generated for short
connection compensation. Refer to
10) for information on generating and using compensation values.
(on page 6-60)
Use the CVU connection compensation value that was generated for load
compensation. Refer to
information on generating and using compensation values.
Model 4200A-SCS Parameter Analyzer Reference Manual
Connection compensation
Connection compensation
Connection compensation
4200A-901-01 Rev. C / February 2017
(on page 4-
(on page 4-
(on page 4-10) for

Advertisement

Table of Contents
loading

Table of Contents