The operation mode determines what type of test is run on the terminal. Selecting the appropriate
mode simplifies configuration options.
A Pulse Sweep and Pulse Step function are identical. However, in a test where two or more PMUs
are used, the sweep for one PMU is performed on each step of the pulse step of the other PMU. A
pulse step requires that at least one channel be configured for Pulse Sweep.
A Pulse Train outputs one or more pulses of the same magnitude and base level.
When there are two or more PMUs in the test system, the
Sweep Master
(on page 6-105) option is
used to select one of them as the master PMU.
Figure 245: Pulse operation modes