Section 6: Clarius
Disable outputs at completion
When a test is complete, you can disable the outputs or leave them enabled.
To set output action on completion:
1. Select the test.
2. Select Configure.
3. In the right pane, select the Test Settings tab.
4. Select Advanced.
5. Select or clear Disable Outputs at Completion.
Current Spot Mean High
Makes current measurements on the amplitude. Available when the Test Mode is set to Pulse I-V.
The next figure shows an example of a spot mean measurement on pulse high (amplitude) and pulse
low (base level). For a more detailed example, refer to
Current Spot Mean Low
Makes current measurements on the base level. Available when the Test Mode is set to Pulse IV.
The next figure shows an example of a spot mean measurement on pulse high (amplitude) and pulse
low (base level). For a more detailed example, refer to
6-72
Model 4200A-SCS Parameter Analyzer Reference Manual
Test Mode (PMU)
Figure 249: Spot mean measurements
Test Mode (PMU)
Figure 250: Spot mean measurements
(on page 6-97).
(on page 6-97).