Section 6: Clarius
Parameters
If you change the Parameters, it will change the Column names, which can cause Formulator
functions to be erased. When using any of the tests or libraries supplied by Keithley Instruments,
leave the measurement option set to Cp-Gp.
You can select the following measurement options:
Z, Theta: Impedance and phase angle (degrees)
R+jX: Resistance and reactance
Cp-Gp: Parallel capacitance and conductance
Cs-Rs: Series capacitance and resistance
Cp-D: Parallel capacitance and dissipation factor
Cs-D: Series capacitance and dissipation factor
Param1 Column Name
The name that is used for parameter 1 in the Analyze sheet.
If you do not define a name, Clarius assigns a name. The assigned name is a combination of the
parameter name and the terminal label.
This cannot be left blank. If it is left blank, Clarius uses the previous value.
If you change the name, it is updated when the test is run. The results from earlier tests that are
available through Run History are not changed to the new name.
Param2 Column Name
The name that is used for parameter 2 in the Analyze sheet.
If you do not define a name, Clarius assigns a name. The assigned name is a combination of the
parameter name and the terminal label.
This cannot be left blank. If it is left blank, Clarius uses the previous value.
If you change the name, it is updated when the test is run. The results from earlier tests that are
available through Run History are not changed to the new name.
Report Test Conditions
If you select the Report Test Conditions option, the DC bias voltage and drive frequency values that
were used for the test are displayed on the Analyze sheet.
When this option is selected, column name fields are available. These are the names of the columns
in the Analyze sheet where this data is reported. You can change the names of the columns by
entering new values for DCV Column Name and Freq Column Name.
6-58
Model 4200A-SCS Parameter Analyzer Reference Manual
4200A-901-01 Rev. C / February 2017