Keithley 4200A-SCS Reference Manual page 321

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Section 6: Clarius
Parameter
Parameters
(on page 6-58)
Param1 Column Name
58)
Param2 Column Name
58)
Report Test Conditions
58)
DCV Column Name
Freq Column Name
Report Status
Compensation Open
Compensation Short
Compensation Load
Compensation Cable Length
page 6-60)
AC Source V
AC Measure I Range
DC Source V
DC Offset
(on page 6-60)
Capacitance Range Estimator
page 6-61)
6-52
Description
Select the type of parameters. When using any of the tests supplied by
Keithley Instruments, leave the measurement option set to Cp-Gp.
(on page 6-
The name that is used for parameter 1 in the Analyze sheet.
(on page 6-
The name that is used for parameter 2 in the Analyze sheet.
(on page 6-
Select this option to display the DC bias and drive frequency values on the
Analyze sheet.
(on page 6-59)
Available if you select Report Test Conditions. Determines the name of the
column that contains the DC bias information in the Analyze sheet where
this data is reported. You can change the name of the column by typing a
new value.
(on page 6-59)
Available if you select Report Test Conditions. Determines the name of the
column that contains the frequency information in the Analyze sheet where
this data is reported. You can change the name of the column by typing a
new value.
(on page 6-59)
When this option is selected, Clarius records measurement status
information when the test executes. A column of the Analyze spreadsheet
displays this information. Hover over a cell to review the information.
(on page 6-59)
Use the CVU connection compensation value that was generated for open
connection compensation. Refer to
10) for information on generating and using compensation values.
(on page 6-59)
Use the CVU connection compensation value that was generated for short
connection compensation. Refer to
10) for information on generating and using compensation values.
(on page 6-60)
Use the CVU connection compensation value that was generated for load
compensation. Refer to
information on generating and using compensation values.
(on
The cable length that was used to generate connection compensation data.
Make sure this cable length is the same as the Cable Length setting in the
Tools > CVU Connection Compensation dialog box. Refer to
compensation
(on page 6-60)
Selects the terminal to use to source AC drive voltage.
(on page 6-60)
The measure range determines the full-scale measurement span that is
applied to the signal.
(on page 6-60)
Selects the terminal to use to source DC drive voltage.
Allows you offset the voltage by up to ±30 V on one terminal. For example,
this allows you to output voltage from 0 V to 60 V instead of −30 V to +30 V.
(on
The Capacitance Range Estimator automatically calculates the maximum
capacitance value based on the parameter settings in the Advanced settings
dialog box.
Model 4200A-SCS Parameter Analyzer Reference Manual
Connection compensation
Connection compensation
Connection compensation
(on page 4-10) for additional information.
4200A-901-01 Rev. C / February 2017
(on page 4-
(on page 4-
(on page 4-10) for
Connection

Advertisement