Keithley 4200A-SCS Reference Manual page 337

Parameter
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Section 6: Clarius
DC Bias - PMU
When DC Bias operation mode is selected, the PMU outputs the DC base voltage. The next figure
shows a representation of a DC voltage waveform.
The DC Bias operation mode generates data that is recorded in the Analyze pane.
The parameters that are specific to this mode are briefly described in the following table. Select the
links to access additional information. Additional parameters are described in
PMU operation modes
the All Parameters pane.
Parameter
Base
(on page 6-71)
Force Range
6-68
Figure 247: DC Bias waveform
(on page 6-69). The parameters are listed in the order in which they appear in
Description
The voltage offset (from 0 V) that is the reference for the pulse amplitude.
(on page 6-71)
The range that is used when sourcing. Select one of the listed ranges. The
source remains on the range that is set. If you are sweeping and a sweep
point exceeds the source range capability, the source outputs the maximum
level for that range. This range must be equal to or greater than the largest
value in the sweep.
Model 4200A-SCS Parameter Analyzer Reference Manual
4200A-901-01 Rev. C / February 2017