10. Perform a one
sweep
(-10 dBm, HP 8722D):
. . . . . . . . . . .
. . . . . . . . . . . : . . : . .
: . . . , . .
Because power meter calibration requires a longer sweep, t@+ you may want
Note
to reduce the number of points before pressing ,r@ C&$ i#@$&? . After the
power meter calibration is finished, return the number of points to its original
11.
To calibrate the R-channel over the IF range, press:
Once completed, the display should read 0 dBm (-10 dBm, HP 8722D).
12.
Make the connections as shown in Figure 3-6, for the one-sweep power meter calibration
over the RF range.

Connections for a One-Sweep Power Meter Calibration for Mixer Measurements

13. 'lb set the frequency offset mode Lo frequency from the analyzer, press:
power meter calibration over the IF frequency range at 0 dBm
Figure 3-6.
L
O