Calibration

Specifying Different Standard for Each Frequency

Specifying Different Standard for Each Frequency

Defining different standard for each frequency band

This section demonstrates the procedure to define a different open standard for each
frequency band, based on the following information.
Table 4-6
Standard
Open
Open
Procedure
Step 1. Press
.
Step 2. Press
.
Cal Kit
Step 3. Select a calibration kit you want to define.
Step 4. Press
Modify Kit
Step 5. Press
Define STDs
Step 6. Select #1, press
Label
screen.
Step 7. Select
as the standard type (
Open
Step 8. Set a necessary standard coefficient.
Step 9. Enter
as the minimum frequency for
1G
Step 10. Enter
as the maximum frequency for
3G
Step 11. Press
.
Return
Step 12. In the same way, select #2, press
that appears on the screen.
Step 13. Select
as the standard type (
Open
Step 14. Set a necessary standard coefficient.
Step 15. Enter
as the minimum frequency for
3G
Step 16. Enter
as the maximum frequency for
6G
Step 17. Press
.
Return
202
Example of standard definition
Label name
Open 3G
Open 6G
.
.
, and enter
using the character input pad that appears on the
Open 3G
STD Type
Min Frequency
Max Frequency
, and enter
Label
STD Type
Min Frequency
Max Frequency
Frequencies to be defined
1 GHz - 3 GHz
3 GHz - 6 GHz
).
.
.
using the character input pad
Open 6G
).
.
.
Chapter 4