Adjustment
Required Adjustment after Replacing Assembly
Required equipment for the synthesizer gain adjustment
None

Virtual Bridge Coefficient Adjustment

This item writes the coefficient factors to the E5071C.
Required equipment for the virtual bridge coefficient adjustment
None
This item is for up to 8.5 GHz Options.

Source Output Power Adjustment

For up to 8.5 GHz Options, the purpose of this procedure is to adjust the
variable attenuator in Level Vernier board, and the gains among Level Vernier,
Distributor and Receiver.
For over 8.5 GHz Options, the purpose of this procedure is to adjust the
characteristic fluctuation of TC709C variable attenuator used in A86 Doubler
Board.
Required equipment for the source output power adjustment
1) Up to 8.5 GHz Options
Description
Power Meter
Power Sensor
2) Over 8.5 GHz Options
Description
Power Meter
Power Sensor
Power Sensor
APC 3.5(m) - N(f)Adapter
APC 3.5(f) -(m) Adapter
Keysight E5071C ENA Network Analyzers
Recommended Model
Keysight E4419A/B
Keysight E9304A Opt H18
Recommended Model
Keysight E4419A/B
Keysight 8482A
Keysight E4413A
Keysight p/n 1250-1811
Keysight p/n 85027-60005