1.5 Demo mode – Repacking andtransport
1.5
Demo mode
Repacking and
1.
Switch the panoramic X-ray unit ON and move it to its packing height by
pressing the
Plus
(XG
– Bite block height = 965 mm (displayed as height on Multipad (XG5/3/3
PPE) or
Easypad (XG
– Bottom edge of slide cover = 702 mm
i
NOTE
The bottom edge of the slide cover must be at the same height as the mark-
ings A in the column.
Start service routine S034, test step 6.
Move the cephalometer to the packing position
(see page 5-165).
Turn the device OFF.
DANGER
PERILOUS SHOCK HAZARD! Prior to disconnecting the line voltage,
you must switch off the power connection and wait 1 minute?.
Unplug any connections between the external units and the DX41 board
or interface board, and disconnect the unit from the power supply.
For information on packing the units and on the packing condition, see the
relevant installation instructions.
transport
i
NOTE
If the unit is installed with a floor stand, its height in-
creases by 30mm. The bite block height value dis-
played on the Multipad (XG5/3/3 PPE) or Easypad
Plus
ready
(XG
/ 3D
) remains the same, however.
/
keys on the Multipad (XG5/3/3 PPE) or Easypad
UP
DOWN
ready
/ 3D
).
Plus
ready
/ 3D
) displayed as height)
A
1 – 13