8.3 Tube Current Verification

8.3 Tube Current Verification
1.
WARNING
The electronics of the X-ray tube assembly are always
connected to line voltage.
Always switch the X-ray unit off and wait until V203 is
no longer illuminated before contacting the test leads.
WARNING
The test leads and measuring instruments used must
have a dielectric strength of at least 1000V!
Be sure to use a battery-powered measuring instru -
ment with shock-hazard-protected sockets.
Use only test leads with shock protection.
NOTICE
The X-ray tube assembly moves during the measurement.
Therefore, be sure to use test leads of sufficient length and
place the measuring instrument in a location where it is firm -
ly seated so that it doesn't fall down.
76
Installation Instructions ORTHOPHOS XG 3
2.
1.
Remove cover (for details see Service Manual).
2.
Loosen the four screws A and remove cover plate B of
the electronics box.
Sirona Dental Systems GmbH
A
B
60 51 416 D 3352
D 3352.031.03.18.02