Checking And Setting Brakes And Spring Force On The Dentist Element - Sirona Intego Service Manual

Hide thumbs Also See for Intego:
Table of Contents

Advertisement

8 Mechanical adjustments
8.5 Checking and setting brakes and spring force on the dentist element
126
8.5
Checking and setting brakes and spring force on
the dentist element
8.5.1
Checking the spring force of the flexible support arm
All instruments are attached to the dentist element and are in the
holders.
1. If a tray holder is attached to the dentist element, attach a tray to the
tray holder.
2. Load the dentist element and tray with typical equipment.
3. Switch the treatment center off.
4. Release the remaining air pressure by actuating the 3-way syringe /
SPRAYVIT E.
 If the dental treatment center is switched off, the dentist element
can swing up or down slightly because the pneumatic brake is not
in operation. This is due to the design and is completely normal.
Sirona Dental Systems GmbH
Service Manual INTEGO / INTEGO pro
64 57 316 D3543
D3543.076.01.04.02 06.2015

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Intego pro

Table of Contents