7.8 Checking the tube data

7 – 14
7.8

Checking the tube data

7.8.1

kV actual value and radiation time

Select service routine S002.3 for the exposure (radiation without rotary
movement, radiation time selectable, see page 5-28) and set the following
parameters:
– Radiation time: 1 s
– kV value: 66 kV
Release radiation and check the kV actual value and radiation time spec-
trometrically, e.g. using a Mult-O-Meter.
If the measured values deviate from the values set above (1 s, 66 kV) by
more than ± 10 %:
Replace the tube assembly (see Page 6-48)
i
NOTE
For the XG 3 / 3 PPE system version, sensor positioning must be performed
until the beam strikes the sensor precisely. If the sensor is positioned correctly,
the corresponding values can be compared with the setpoint values.
7.8.2

Checking the tube current

DANGER
The electronics of the X-ray tube assembly are always connected to line
voltage.
The device must be switched off and, after switching off the unit, wait at
least 4 minutes (LED V203 must not be on) before you ...
– remove the cover from the tube assembly
– Attach measuring wires to the DX6 board, or
– detach the jumper from the mA measuring socket
WARNING
The test leads and measuring instruments used must have a dielectric
strength of at least 1000V!
Be sure to use a battery-powered measuring instrument with shock-hazard-
protected sockets.
Use only test leads with shock protection.