Assessing X-Ray Exposures; Checking The Tube Data - Sirona ORTHOPHOS XG 5 Maintenance Manual

Hide thumbs Also See for ORTHOPHOS XG 5:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
4.7

Assessing X-ray exposures

Ask to be shown the most recent X-ray exposures taken by the dentist.
Assessing X-ray exposures
Maintenance interval
(after years)
Date
(enter month/year)
Unexposed surrounding border
present?
See Service Manual
Exposure sharp?
Density of X-ray image OK ?
4.8

Checking the tube data

Checking the tube data
Maintenance interval
(after years)
Date
(enter month/year)
Actual kV/mA values OK?
See Service Manual
Fan working?
Does the temperature sensor in
the tube assembly indicate a
value between ca. 18° and 65°C?
59 87 685 D 3352
D 3352.105.01.11.02
4
7
10
If no  Correct fault.
Answer questions with yes () or no (−)
4
7
10
If no  Correct fault.
4 Maintenance by the service engineer
Answer questions with yes () or no (−)
12
14
12
14
23