10 Maintenance

10.9 Checking the device leakage current

A
382
10.9
Checking the device leakage current
Checking the device leakage current
DANGER
Potentially lethal shock hazard!
It is essential to switch the unit off and to wait at least one more 1 minute
before beginning the check!
A high resistance measuring voltage source at line frequency and a
measuring circuit compliant with the requirements of IEC 60601 are
required. Complete test units, e.g. the "Bender tester", fulfill these
requirements.
Measuring setup for testing the device leakage current
B
L
N
PE
A
A
Power cable to the unit
B
Voltage source: 200-240V, 50Hz/60Hz
D
Voltmeter
Sirona Dental Systems GmbH
Service Manual GALILEOS
MD
1kOhm
0,015µF
10kOhm
V
D
D3437.076.01.15.02 06.2012
B
61 25 665 D3437