10 Maintenance
10.10 Checking the tube data
Preparing the measurement
D
458
10.10
Checking the tube data
10.10.1
Checking the tube voltage
10.10.1.1
Checking the tube voltage characteristic during exposure (tube
assembly 1.0)
NOTICE
Damage to the measuring unit
The ring assembly and the tube assembly move during the
measurement.
Make sure that the measuring wires are sufficiently long to allow for the
ring movement and that the measuring unit is in a secure position so that
it will not fall down.
1. DANGER! After switching off the unit, wait at least 4 minutes (LED
V203 must not be on) before removing the cover at the tube
assembly!
Switch the unit off (see Operating Instructions).
2. Loosen the 3 screws (D) and remove the "front tube assembly" cover
D
(E).
D
E
3. Loosen the 4 screws (F) and remove the cover plate (G).
F
G
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
63 03 510 D3352