5 Messages

5.5 List of error messages

Error code
Description
E6 06 01
General error during module
initialization
Error code
Description
E6 06 02
Invalid system data or uninitialized
module storage data
Error code
Description
E6 06 03
Invalid commanding of control data,
CAN bus error
This error may also occur in
connection with other causal error
messages! Please also observe the
causal error message! It appears
only after you acknowledge the first
error message.
Error code
Description
E6 06 04
Data transfer error or dialog error to
module (master side)
Error code
Description
E6 06 05
Data transfer error or dialog error to
bootloader of module
Only occurs in connection with
software update.
88
5.5
List of error messages
List of error messages
In the following table, the error codes are sorted by the location or function
unit where the error has occurred. For enhanced clarity, the
corresponding ID in the error code is printed in bold type.
5.5.1

Location 06: Tube assembly/DX6

Location 06: Tube assembly/DX6
Actions required
● Please contact the Sirona Customer
Service Center (CSC) to find out
whether a bug fix by means of a
software update is possible and
perform such an update if necessary.
If the error occurs repeatedly ...
● Replace the tube assembly.
E6 06 02 GALILEOS
Actions required
● Please contact the Sirona Customer
Service Center (CSC) to find out
whether a bug fix by means of a
software update is possible and
perform such an update if necessary.
E6 06 03
Actions required
● Check the CAN bus.
● Please contact the Sirona Customer
Service Center (CSC) to find out
whether a bug fix by means of a
software update is possible and
perform such an update if necessary.
E6 06 04
Actions required
● Check the CAN bus.
● Please contact the Sirona Customer
Service Center (CSC) to find out
whether a bug fix by means of a
software update is possible and
perform such an update if necessary.
E6 06 05
Actions required
● Repeat the software update.
● Check the CAN bus.
If the error occurs repeatedly or the module
is no longer addressable ...
● Replace the tube assembly.
Sirona Dental Systems GmbH
Service Manual GALILEOS
see
S. [ → 61]
S. [ → 310]
see
S. [ → 61]
see
S. [ → 128]
S. [ → 61]
see
S. [ → 128]
S. [ → 61]
see
S. [ → 61]
S. [ → 128]
S. [ → 310]
61 25 665 D3437
D3437.076.01.15.02 06.2012

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents