Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
63 03 510 D3352
D3352.076.03.11.02 09.2016
What has to be done after replacing the 2D sensor (Pan)?
9.13.3.1
IMPORTANT
Use the most current software version available on the ORTHOPHOS
XG3D CD.
1. Switch the unit on.
2. Perform a software update to the current unit software version [ → 62].
All modules are updated in accordance with the configuration.
3. Switch the unit off.
Wait approx. 1 minute. Then switch the unit on again.
4. Perform a complete unit adjustment or calibration [ → 184].
Only for Germany, Austria, Switzerland and France:
5.
Perform an acceptance check without consulting an expert.
9 Repair
9.13 Combisensor
423