9 Repair
9.2 Height adjustment motor (M1_4)/spindle
300
6. Lay the motor cable in the cable harness (V) on the rear of the unit
and secure in position with the cable clamps.
7. Route the cable into the arm.
IMPORTANT: The green mark must lie in the recess (X).
8. Plug connector X402 (W) into board DX1.
9.2.5
What has to be done after replacing the height adjustment
motor (M1_4) or the spindle?
What has to be done after replacing the height adjustment motor (M1_4) or the spindle?
1. After inserting the new spindle above and below the height
adjustment motor, grease it thoroughly with Chesterton 622.
2. Use the Up/Down keys on the control panel to check the function of
the height adjustment motor.
3. Reset the travel height.