Removing the covers
Removing the defective laser module
Installing the new laser module
59 38 399 D3352
D3352.076.01.26.02 09.2016

6.9 Replacing/adjusting the MS light localizer laser module (PAN)

6.9
Replacing/adjusting the MS light localizer
laser module (PAN)
DX1
1.
Remove the covers (see also section 1.13):
– Top arm
Pull the cables of the laser modules (FH+MS) off connectors X807 (FH)
and X811 (MS) on board DX1.
Detach the cover (see also section 1.13):
– Front slide cover, carefully pulling it together with the FH light localizer
off toward the front.
Detach the cable from the cable holder.
1.
Loosen the two screws A and carefully remove the laser module (MS) to-
ward the front.
Insert the new laser module (MS) in the holder and tighten the screws A
loosely.
i
NOTE
Make sure that the lines in the laser module run horizontally.
Plug the cable of the new MS laser module into connector X811 on board
DX1 and run the cable in the cable holder.
i
NOTE
Before reattaching the front slide cover, you must adjust the MS light localizer.