Adjusting The Fh Double Laser Light Localizer (Pan) - Sirona Orthophos XG 3D Ceph Service Manual

Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
A
B
T
R
63 03 510 D3352
D3352.076.03.11.02 09.2016
5. Loosen the screws (B) and carefully pull the laser module (C) out of
the holder toward the front.
6. Insert the new laser module (FH) in the holder and retighten screws
(B).
IMPORTANT
Make sure that the lines in the laser module run vertically.
7. Screw the holder securely back onto the cover with screws (A) and
reattach the strain reliefs.
8. Set the "Front slide" cover down on the "Top support" cover.
9. Plug the cables of the FH laser modules into the sockets of cable L14.
10. Next, adjust the double laser light localizer [ → 393].
11. Reattach the "Front slide" cover [ → 41].

Adjusting the FH double laser light localizer (PAN)

9.7.1.2
The "Front slide" cover must have been removed [ → 41].
1. Switch the unit on via the switch (A) (see also Operating Instructions).
The X-Ray radiation indicator (B) lights up briefly.
After approx. 2 seconds, the green LED (C) in the upper part of
the control panel lights up. This LED remains lit as long as the
unit is on.
The start screen appears on the touchscreen of the Easypad and
the system's self-adjustment routine starts running (for approx. 1
minute). The diaphragm moves into position. Forehead and
C
temple supports on the panoramic unit are moved to the home
position.
On completion of the self-adjustment routine, the main menu
appears on the touchscreen. Help message H301 prompts you to
move the unit into the starting position.
2. Touch the 3D symbol at the top of the touchscreen.
The exposure program VOL1 is displayed.
3. Press the R key.
The diaphragm and the sensor move into the starting position for
volume exposures.
9 Repair
9.7 Laser light localizers
393