Checking The Laser - Sirona Galileos Maintenance Instructions Manual

Hide thumbs Also See for Galileos:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Maintenance Instructions GALILEOS
d.

10 Checking the laser

1.
A
1.
Place the bite block A in the bite block holder for the
unit (see User Manual).
2.
Check the laser:
Switch on the light localizer with key B on the
Easypad. The Lasers generate a red line.
WARNING
Class 2 radiation is emitted during installation.
Always keep eyes a minimum distance of 100mm away
from the laser. Do not stare into the beam.
SAG laser (vertical line)
The laser beam must be visible in the center of the bite
block as a vertical line.
i
NOTE
To adjust the lasers, see Service Manual
No controls are available to adjust the laser power.
61 16 813 D3437
D3437.103.01.04.02
2.
Label
10 Checking the laser
9.9 Geometric calibration
B
33