Sirona Dental Systems GmbH
Installation Instructions ORTHOPHOS XG 3
8.5 Checking the laser
1.
1.
Insert the forehead and temple supports.
2.
Check the laser:
•
Fasten a piece of white cardboard between the temple
supports.
•
Swivel the mirror by pressing the left recess A on the
touch bar.
Switch on the light localizer with key B on the Easy -
•
pad. The Lasers generate a red line.
WARNING
Class 1 radiation is emitted during installation.
Always keep eyes a minimum distance of 100 mm away
from the laser. Do not stare into the beam.
SAG laser (vertical line)
The light beam must strike the center of the head sup -
port in a vertical direction.
FH laser (horizontal line)
The light beam must strike the FH line of the sheet of
cardboard attached to the head support. It must be pos -
sible to displace the light beam vertically using slide C.
Unwanted movement of the FH light localizer is not per -
mitted.
IMPORTANT
To adjust the lasers, see Maintenance Instructions.
No controls are available to adjust the laser power.
60 51 416 D 3352
D 3352.031.03.18.02
for USA/Canada only
2.
C
A
Label
8 Startup for USA/Canada only
8.5 Checking the laser for USA/Canada only