Maintenance Instructions
1.
32°
Horizontal beam
Checking the Laser Light Beam
Pressing the height adjustment button ↑ always automatically switches on the laser light localizers for one minute.
•
•
Radiation of class 1 appears during tests.
Keep always a minimum distance between eyes and laser of 100mm.
•
No stray radiation shall be emitted by defective optics or covers.
The laser must generate a line.
•
No controls are available to adjust the level of laser radiation.
1. Horizontal laser light beam FH.
The light beam must meet the FH line of a sheet of paper which is fixed in the head support. It must be
moveable in the vertical direction by moving the laser module at the cover of ORTHOPHOS 3. Unwan-
ted movement of the FH-light localizer is not permitted.
2. Vertical laser light beam.
The light beam must meet the center of the head support in a vertical direction.