Checking The Laser Light Beam - Sirona Orthophos 3 Maintenance Manual

Hide thumbs Also See for Orthophos 3:
Table of Contents

Advertisement

Maintenance Instructions
1.
32°
Horizontal beam

Checking the Laser Light Beam

Pressing the height adjustment button ↑ always automatically switches on the laser light localizers for one minute.
Radiation of class 1 appears during tests.
Keep always a minimum distance between eyes and laser of 100mm.
No stray radiation shall be emitted by defective optics or covers.
The laser must generate a line.
No controls are available to adjust the level of laser radiation.
1. Horizontal laser light beam FH.
The light beam must meet the FH line of a sheet of paper which is fixed in the head support. It must be
moveable in the vertical direction by moving the laser module at the cover of ORTHOPHOS 3. Unwan-
ted movement of the FH-light localizer is not permitted.
2. Vertical laser light beam.
The light beam must meet the center of the head support in a vertical direction.
20
60°
Sirona Dental Systems GmbH
2.
Vertical beam
59 01 603 D3285
D3285.103.04.02.02