12 Basic Workstation Settings with Mode Key
A
CUP FILL ON S
YES − NO
A
MAX. LIGHT VOLT
3.9 V WARNING
A
TAKE INSTRUMENT
A
MAX DRIVE INTEN.
34
A
TAKE INSTRUMENT
A
START INSTRUMENT
A
RPM
Sirona Dental Systems GmbH
12.8
Mode: CUP FILL ON S
YES - NO
Filling of tumbler in chair program S
The tumbler is filled when the treatment
chair is in mouth rinsing position S (See
page 19).
12.9

Mode: MAX LIGHT VOLT

Instrument light, voltage setting for
instruments.
First a message will appear prompting you
to TAKE an INSTRUMENT.
Then the current voltage setting of the
withdrawn instrument is displayed.
Using the – / + keys, you can then set the
instrument light voltage from 3.0 to 5.6 V in
steps of 0.1 V.
Starting from 3.9 V, WARNING appears
on the display.
After the instrument has been returned to
its hanger, please set the next instrument.
Do not continue on in the Mode menu
before the instrument light voltage has
been set for all instruments.
12.10

Mode: MAX DRIVE INTEN.

For SIRONA motor and turbine
First a message will appear prompting you
to remove an instrument.
Then the instrument must be started.
The maximum permissible motor speed in
RPM for the SIRONA motor and the
maximum permissible air pressure in bar
(or PSI) for the turbine can be set with
+
the
/
keys.
Once an instrument has been returned to
its hanger, please set the next instrument.
bar
Do not continue on in the Mode menu
before the maximum power has been set
for all instruments.