Adjusting the MS laser light localizer (PAN)

9.7.3.2
The "Front slide" cover must have been removed [ → 41].
1. Insert the forehead and temple supports (see also Operating
Instructions).
2. Switch the unit on via the switch (A) (see also Operating Instructions).
The X-Ray radiation indicator (B) lights up briefly.
After approx. 2 seconds, the green LED (C) in the upper part of
the control panel lights up. This LED remains lit as long as the
unit is on.
The start screen appears on the touchscreen of the Easypad and
the system's self-adjustment routine starts running (for approx. 1
minute). The diaphragm moves into position. Forehead and
C
temple supports on the panoramic unit are moved to the home
position.
On completion of the self-adjustment routine, the main menu
appears on the touchscreen. Help message H301 prompts you to
move the unit into the starting position.
3. Touch the PAN symbol at the top of the touchscreen.
The program group PAN is selected.
4. Press the R key.
The diaphragm and the sensor move into the starting position for
panoramic exposures.
5. Press the light localizer key on the control panel.
The light localizers are switched on.
The vertical laser beam must be displayed in the center of the
forehead support and the bite block holder.
9 Repair
9.7 Laser light localizers
399