What Has To Be Done After Replacing The Control Panel; Laying Cables For Control Panel Replacement - Sirona Orthophos XG 3D Ceph Service Manual

Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
63 03 510 D3352
D3352.076.03.11.02 09.2016

What has to be done after replacing the control panel?

9.6.2.1
IMPORTANT
Use the most current software version available on the ORTHOPHOS
XG3D CD.
1. Switch the unit on.
2. Perform a software update to the current unit software version [ → 62].
3. Switch the unit off.
Wait approx. 1 minute. Then switch the unit on again.
4. Check the control panel and the display elements for correct
functioning: When the unit is switched on, all of the display elements
must light up briefly.
9.6.3

Laying cables for control panel replacement

A
Strain relief
9 Repair
9.6 Control panel
391