Replacing The Control Panel - Sirona ORTHOPHOS XG 3Dready Service Manual

Table of Contents

Advertisement

Tab 6
4.
C
Removing the user interface
Removing the defective control panel
Integrating a newcontrol panel
59 38 399 D3352
D3352.076.01.26.02 09.2016
6.7

Replacing the control panel

2.
B
Remove the user interface (see page 6-26).
1.
Swing folding mechanism A for the user interface all the way up. This
makes it easier to remove the cover.
2.
Press into slit B of the housing cover with a screwdriver (do not pry!) and
remove the cover.
3.
Detach cables L9 and L10 from the strain reliefs.
4.
Rotate the control panel to the center position and loosen the four screws
C.
Pull the cables out of the control panel and remove the panel.
To install the control panel and the user interface, follow the procedure for
removing them in reverse order.
i
NOTE
Don't forget to plug connectors X102 and X103 back into board
DX7 / DX71, lay the cables in their original positions and reattach the strain re-
liefs (see "Laying cables when replacing the control panel" on page 6-30).
ATTENTION: What to do after replacement?
- Check the user interface for correct functioning as well as the
Function of the display elements. (After the unit is switched ON
all of the display elements must light up briefly!)
– No further action is required.

6.7 Replacing the control panel

1.
A
6 – 29