7 Initial startup

7.3 Switching the units ON

7.3 Switching the units ON
1.
WARNING
When performing the following test, be sure to observe
the radiation protection regulations applicable in your
country (see operating instructions).
It is prohibited for any person to be positioned in the
unit when it is switched on.
NOTICE
If the room height is less than 2.27 m (89 3/8") (2.30 m (90
1/2") with floor stand), you must limit the maximum travel
height of the unit (see section 11.1.7).
68
?
?
Installation Instructions ORTHOPHOS XG 3
2.
1.
Switch the unit ON.
– All LEDs and the LED display on the Multipad light
up briefly.
– The green LED at the top of the Multipad
permanently as long as the unit is ON.
– The initialization status is visualized by a progress
indicator while the unit executes a self-adjustment
routine (approx. ½ min.). At the same time, the
rotating element rotates briefly clockwise and coun -
terclockwise and the diaphragm is positioned. The
forehead and temple supports on the panoramic
unit open and close and then stop moving in fully
opened position.
– Once the self-adjustment routine is completed, help
message H301 prompts you to move the unit to the
starting position.
2.
Press the R key to move the unit back to the starting
position.
Switch the PC ON and start SIDEXIS XG.