6 Installation and startup
6.4 Initial startup
A
28
Operating Instructions for the Acquisition Unit Sirona Dental CAD/CAM System
6.4.6

Using a trackball

1. Turn the collar (A) counterclockwise and remove it.
Using a trackball
2. Insert the ball supplied.
3. Lay the collar (A) into position and turn it clockwise until it snaps into
place.
6.4.7

Changing from right-handed to left-handed operation

Changing from right-handed to left-handed operation
In the factory default setting, the left button trackball button corresponds
to a foot control entry. If you would like to change this assignment to the
right trackball button, your CEREC service technician can do this for you.
6.4.8

Switching the units on

Note on condensate AU
Switching the units on
NOTICE
Do not put the unit into operation at low temperatures!
If you move the unit to the operating site from a cold environment,
condensation may form and result in a short circuit.
Install the unit at room temperature.
➢ Wait until the unit has reached room temperature and is absolutely
dry (for at least one hour)
 The unit is dry and can be put into operation.
Description
If the acquisition unit is switched on at the main switch, it can be activated
with the ON button. The monitor is switched on and off automatically (if it
was switched on before the acquisition unit was switched off).
You can switch the monitor on and off with the monitor ON/OFF switch.
Sirona Dental Systems GmbH
63 33 582 D3492
D3492.201.01.09.23 09.2011