5.18 Service routine S033 (not for XG 3 / 3 PPE)

Function S033
Test step 10
Selecting service routine S033
Selecting test step 10
5 – 152
5.18
Test of CEPH image path without SIDEXIS
Test of the CEPH image path without SIDEXIS
Select the Service menu (see page 5-12).
1.
Use the arrow keys to select service routine S033.
Plus
XG
Easypad touchscreen
1
-
2
3
5.18.1
1.
Select test step 10 in selection field 2 with the arrow keys and confirm
your selection by pressing the Service key
Plus
XG
Easypad touchscreen
1
-
-
2
3
After test step 10 is selected, selection field 1 displays the current selection
for the synchronized readout sequence.

Service routine S033 (not for XG 3 / 3 PPE)

Test of CEPH image path without SIDEXIS
ready
/ 3D
:
+
S033

S033: Test step 10

ready
/ 3D
:
+
S033
S33
10
+
Tab 5
XG 5: Multipad
S033_______10___
1
2
.
XG 5: Multipad
S033_______10___
1
2
59 38 399 D3352
D3352.076.01.26.02 09.2016
3
3