9.13 Cable
Prepare the cable exchange
Replacing cables
364
Cable replacement (L3, L5, L6, and L15)/Laying the cable/
9.13.2.2

corrugated tube at the rotation unit

Cable replacement (L3, L5, L6, and L15)/Laying the cable/corrugated tube at the rotation unit
NOTICE
The connectors and cables must be protected by inserting them in the
fabric tube (A) supplied with the cables.
1. Remove the covers.
2. Pull the connectors off board DX6.
1. For L3: Remove the connector of the cable with tool W1.
2. Remove the corrugated tube and the spiral spring (B) from the cable
loom.
3. Remove the defective cable and run the new cable up to the rotary
ring in the original position.
4. Bunch the cables together again to form a loom.
5. Fasten the defective cable to the loom and use it as a pull wire to pull
the loom through the fabric tube (A).
6. Pull the fabric tube over the connector and as far over the cable loom
as possible.
7. Use the pull wire to pull the fabric tube into the spiral spring (B).
Sirona Dental Systems GmbH
Service Manual GALILEOS
61 25 665 D3437
D3437.076.01.15.02 06.2012