13 Adjusting and calibrating the unit
13.2 Adjustment and calibration via the "Unit adjustment/calibration" menu
228

Ceph - Fixed point of rotation

13.2.1.5
13.2.1.5.1
Automatic adjustment: Ceph - Fixed point of rotation (standard)
NOTICE
Adjustment 1
Risk of damage to unit
It is essential that the OP XG needle phantom is removed from the bite
block holder on the device before a ceph exposure is taken; otherwise,
the combisensor could collide with the phantom.
The OP XG needle phantom is
unit.
The OP XG ceph adjustment phantom is
unit.
1. In the structure tree, under
Fixed point of rotation"
 The
action area. The menu supports precision adjustment and coarse
adjustment (precision adjustment is pre-selected). Perform a
precision adjustment first. In most cases, previous coarse
adjustment is not necessary.
2. Make SIDEXIS XG ready for exposure. [ → 206]
3. Take an exposure (80 kV/14 mA; 0.6 s) [ → 206].
 SIDEXIS XG calculates the adjustment values S1, S2, and S3
automatically from the exposure and enters them in the text
boxes of the
Installation Manual ORTHOPHOS XG 3D / Ceph
not
"2D Adjustment"
element (S010.5).
"Ceph - Fixed point of rotation"
"Ceph - Fixed point of rotation"
Sirona Dental Systems GmbH
in the pan bite block holder on the
not
in the pan slot on the
, click on the
"Ceph -
menu is displayed in the
menu.
63 03 452 D3352
D3352.031.05.08.02 10.2015