Service Routine S014; S014: Test Step 1 - Sirona ORTHOPHOS XG 3Dready Service Manual

Table of Contents

Advertisement

S014: Rotation motor service

5.11 Service routine S014

Function S014
Test step 1
Test step 2
Test step 3
Test step 4
Selecting service routine S014
Selecting test step 1
5 – 74
5.11
Rotation motor service
Travel of rotational drive to the Pan home position
Travel of rotational drive to the ceph home position
Free travel of rotational drive
Display of light barrier signals of rotational drive
Select the Service menu (see page 5-12).
1.
Use the arrow keys to select service routine S014.
Plus
XG
Easypad touchscreen
1
-
-
2
3
5.11.1
This test step is no longer offered in all software versions.
This test step realizes "isolated" travel of the rotation motor to the pan home
position. Actuators 1 and 2 remain in their current positions.
1.
Select test step 1 in selection field 2 with the arrow keys and confirm your
selection by pressing the service key
Plus
XG
Easypad touchscreen
1
-
-
2
3
After the test step is selected, selection field 1 displays the current status:

Service routine S014

Rotation motor service
ready
/ 3D
:
+
S014
1
+

S014: Test step 1

ready
/ 3D
:
+
S014
+
1
Tab 5
XG 5 / 3 / 3 PPE: Multipad
S014_______1____
1
2
.
XG 5 / 3 / 3 PPE: Multipad
S014_______1____
1
2
59 38 399 D3352
D3352.076.01.26.02 09.2016
3
3

Hide quick links: