2 Safety information
2.4 Modifications and extensions of the system
8
Operating instructions 22'' Monitor AC, REF 64 65 228 SIVISION digital
2.4

Modifications and extensions of the system

Modifications to this system which might affect the safety of the system
owner, patients or other persons are prohibited by law.
For reasons of product safety, this product may be operated only with
original Sirona accessories or third-party accessories expressly approved
by Sirona. The user assumes the risk of using non-approved accessories.
If any devices not approved by Sirona are connected, they must comply
with the applicable standards, e.g.:
● IEC 60950-1 for information technology equipment (e.g. PC) and
● IEC 60601-1 for medical electrical equipment.
The loudspeaker port of the monitor may be connected only to a device
that complies with IEC 60950 (e.g. a PC) or
IEC 60601-1. Under no circumstances may it be connected to a stereo
system, or similar, for example.
If a system is created during the installation process, the requirements of
IEC 60601-1, 3rd edition, must be fulfilled. The system author assumes
responsibility for conformity of the system with 93/42/EEC Directive.
2.5

Danger of electric shock

To prevent the danger of fire or of electric shock, the product must not be
exposed to rain or moisture.
Please do not open or dismantle the unit, as otherwise there will be a risk
of electric shock.
2.6

Electromagnetic compatibility

Medical electrical equipment is subject to special precautionary
measures with regard to electromagnetic compatibility (EMC). It must be
installed and operated in compliance with the specifications provided for
the respective treatment center in the document "Installation
Requirements".
Portable and mobile RF communications equipment may interfere with
medical electrical equipment. Therefore, the use of mobile wireless
phones in medical office or hospital environments must be prohibited.
To ensure safe operation from the standpoint of EMC, please observe the
chapter entitled "Electromagnetic compatibility" in the document
"Installation Requirements".
2.7

Dismantling/Installation

When dismantling and reinstalling the unit, proceed according to the
Follow the installation instructions
installation instructions for new installation in order to guarantee its proper
functioning and stability.
This unit must not be used in areas subject to explosion hazards.
Explosion-hazard areas
Sirona Dental Systems GmbH
64 65 228 D3637
D3637.201.01.01.02 15.11.13