Menu level 1
Menu level 2
History
Cycles
Errorbuffer
Service
About

3.8 Resetting the menu system

Should you need to reset the menu – back to factory settings:
1. The display must state "Ready 134C"
2. Press the "Wrapped" button and keep it down as you also press "Menu/Enter"
Your menu is now reset with the following features:
0
Program:
134
C
Printer:
Off
Language:
UK
Oil Expel:
1
Rinse cycles:
2
Expel:
0
60 53 362 D3472
D3472.201.01.09.02
DAC UNIVERSAL – USER MANUAL
729050 - Version 16 S
Menu level 3
Result/action
Displays number of cycles run in total
Displays last machine errors w/Up & Down buttons
Displays when the machine is due for the Next Service
– listed both as number of cycles left until next service
and as the date (M-D-Y) for the next service.
Displays machine serial no. and software version