Resetting Adjustment/Calibration - Sirona Orthophos XG 3D Ceph Service Manual

Table of Contents

Advertisement

7 Adjusting and calibrating the unit
7.2 Adjustment and calibration via the "Unit adjustment/calibration" menu
Reset panoramic adjustment
Reset ceph adjustment
242
7.2.4

Resetting adjustment/calibration

In an absolute emergency, the settings made for adjustment and
calibration can be reset in full or in part to the factory settings and/or
modified manually in the
CAUTION
You must make a note of the values displayed in the text boxes of the
menus before changing them. If necessary, you can then reset the
adjustment values to the factory settings.
Contact the Sirona Customer Service Center for more information (or to
enable the menu).
1. To reset panoramic adjustment, click on the
adjustment"
adjustment"
2. To reset the adjustment values, click on the corresponding button.
1. To reset ceph adjustment, click on the
element in the structure tree under
2. To reset the adjustment values, click the corresponding button.
Service Manual ORTHOPHOS XG 3D / Ceph
"Adjustment/calibration"
element in the structure tree under
.
"2D Reset adjustment"
Sirona Dental Systems GmbH
menu.
"Panorama Reset
"2D Reset
"Ceph - Reset adjustment"
.
63 03 510 D3352