Test Phantoms For Adjustment And Calibration; Op Xg Needle Phantom For Panoramic Adjustment - Sirona Orthophos XG 3D Ceph Service Manual

Table of Contents

Advertisement

7 Adjusting and calibrating the unit
7.1 General information about unit adjustment and calibration
198
7.1.6

Test phantoms for adjustment and calibration

OP XG needle phantom for panoramic adjustment

7.1.6.1
NOTICE
Risk of damage to unit
Make absolutely sure that the OP XG needle phantom is inserted in the
bite block holder of the unit in such a way that the front needle (B) of the
phantom points
figure). Otherwise, this can result in a collision between the combisensor
and the OP XG needle phantom and thereby damage the unit.
NOTICE
Warning for XG 5 / 3D / 3D ready ceph exposures
Risk of damage to unit
It is essential that the needle phantom is removed from the bite block
holder on the device
phantom could collide with the sensor.
For panoramic sensor and pan symmetry panorama adjustments, the OP
XG needle phantom (A) must be inserted into the pan bite block holder of
the unit. For pan diaphragm panorama adjustment, the needle phantom
must be taken out of the bite block holder.
Service Manual ORTHOPHOS XG 3D / Ceph
away from the unit in the direction of the patient
before
a ceph exposure is taken; otherwise, the
Sirona Dental Systems GmbH
(see
B
A
63 03 510 D3352
D3352.076.03.11.02 09.2016