Test Phantoms For Adjustment And Calibration; Op Xg Needle Phantom For Panoramic Adjustment - Sirona Orthophos XG 3D/Ceph Installation Manual

Table of Contents

Advertisement

Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS XG 3D / Ceph
63 03 452 D3352
D3352.031.05.08.02 10.2015
13.1.6

Test phantoms for adjustment and calibration

OP XG needle phantom for panoramic adjustment

13.1.6.1
NOTICE
Risk of damage to unit
Make absolutely sure that the OP XG needle phantom is inserted in the
bite block holder of the unit in such a way that the front needle (B) of the
phantom points
figure). Otherwise, this can result in a collision between the combisensor
and the OP XG needle phantom and thereby damage the unit.
NOTICE
Warning for XG 5 / 3D / 3D ready ceph exposures
Risk of damage to unit
It is essential that the needle phantom is removed from the bite block
holder on the device
phantom could collide with the sensor.
For panoramic sensor and pan symmetry panorama adjustments, the OP
XG needle phantom (A) must be inserted into the pan bite block holder of
the unit. For pan diaphragm panorama adjustment, the needle phantom
must be taken out of the bite block holder.
13 Adjusting and calibrating the unit
13.1 General information about unit adjustment and calibration
away from the unit in the direction of the patient
before
a ceph exposure is taken; otherwise, the
(see
B
A
207