Removing/inserting the sensor

NOTICE
Risk of damage to the sensor
When removing the sensor, it can be damaged by impact or if dropped.
The sensor contains an integrated vibration sensor to detect impacts or
falls. If the vibration sensor has triggered, guarantee claims become
void.
➢ Do not drop the sensor under any circumstances!
Electrostatic charge
NOTICE
Electrostatic charges from persons are discharged on the unit.
Electrical components of the unit are destroyed.
➢ Do not touch any electrical components or unprotected plug
contacts.
➢ Discharge yourself by touching a conductive grounded object.
Removing/inserting the sensor
Removing the sensor
1. Hold the sensor firmly.
2. Press the button all the way in and hold it.
The sensor is released from the fastening.
3. Pull the sensor downwards out of the guide.
Inserting the sensor
1. Hold the sensor firmly.
2. Using both guide pins, insert the sensor into the guide sleeves on the
unit and push until it reaches a stop.
The sensor engages in the X-ray unit.
4 Installation and start-up
4.3 Removing/inserting the sensor
31