1.7 Software update
1 – 36
1.7.5
Software update of ORTHOPHOS XG
version V02.22 to V02.30 or higher
i
NOTE
If the sensors previously used are to be used in the future as well, plug them
into the slots on the panoramic unit and/or cephalometer before you begin the
update.
1.
Switch the unit ON.
2.
Perform the software update to version V02.30 or higher
(automatic update) as described on page 1-29.
3.
Turn the device OFF.
Wait for approx. 1 minute. Then switch the unit ON again.
4.
Perform a ceph adjustment if a cephalometer is installed (see chapter
4.5).
5.
Check to see which Sidexis software version is installed:
– If SIDEXIS software version V01.55 is installed, no update is required.
– If SIDEXIS software version <V01.55 or V01.6 is installed, then update
the SIDEXIS software to version V02.4 or higher.
– If a SIDEXIS software version between V02.0 and V02.3 is installed,
then SIRONA recommends updating the SIDEXIS software to version
V02.4 or higher.
CAUTION
When doing so, observe the system requirements for SIDEXIS V02.4. If a
V01.55 software version was previously installed, the database also must be
adapted accordingly. If this is not possible, please contact the CSC.
The process is completed.