Software Update Orthophos Xg Plus From Version V02.20 To V02.30 Or Higher - Sirona ORTHOPHOS XG 3Dready Service Manual

Table of Contents

Advertisement

Tab 1
E6 15 04
59 38 399 D3352
D3352.076.01.26.02 09.2016
1.7.4
Software update ORTHOPHOS XG
from version V02.20 to V02.30 or higher
i
NOTE
If the sensors previously used are to be used in the future as well, plug them
into the slots on the panoramic unit and/or cephalometer before you begin the
update.
1.
Switch the unit ON.
2.
Perform the software update to version V02.25 or higher
(automatic update) as described on page 1-29.
3.
Turn the device OFF.
Wait for approx. 1 minute. Then switch the unit ON again.
Error message E6 15 04 (undefined activation data) is displayed.
4.
Acknowledge the error message with the R key
R
5.
Open the Service menu (see page 5-12).
6.
Start service routine S008.3 and confirm the unit
serial number (see page 5-55).
i
NOTE
The unit serial number is located on the rating plate of the unit.
CAUTION
In case of a wrong serial number, cancel the update and contact the SIRONA
Customer Service Center!
7.
Start service routine S008.4 and enable the function activation. Exit the
Service menu again.
8.
Turn the device OFF.
Wait for approx. 1 minute. Then switch the unit ON again.
9.
Perform the complete system adjustment (see chapter 4):
– PAN adjustment
– For TSA units: TSA adjustment
– CEPH adjustment if cephalometer is installed
10. Check to see which Sidexis software version is installed:
– If SIDEXIS software version V01.55 is installed, no update is required.
– If SIDEXIS software version <V01.55 or V01.6 is installed, then update
the SIDEXIS software to version V02.4 or higher.
– If a SIDEXIS software version between V02.0 and V02.3 is installed,
then SIRONA recommends updating the SIDEXIS software to version
V02.4 or higher.
CAUTION
When doing so, observe the system requirements for SIDEXIS V02.4. If a
V01.55 software version was previously installed, the database also must be
adapted accordingly. If this is not possible, please contact the CSC.
The process is completed.
1.7 Software update
Plus
.
R
1 – 35

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents