Command*
Linear staircase sweep:
*RST
:SOUR:VOLT 0
:SOUR:DEL 0.1
:SOUR:SWE:RANG BEST
:SOUR:VOLT:MODE SWE
:SOUR:SWE:SPAC LIN
:SOUR:VOLT:STAR 1
:SOUR:VOLT:STOP 10
:SOUR:VOLT:STEP 1
:TRIG:COUN 10
:OUTP ON
:READ?
Logarithmic staircase sweep:
*RST
:SOUR:VOLT 0
:SOUR:DEL 0.1
:SOUR:SWE:RANG BEST
:SOUR:VOLT:MODE SWE
:SOUR:SWE:SPAC LOG
:SOUR:VOLT:STAR 1
:SOUR:VOLT:STOP 10
:SOUR:SWE:POIN 5
:TRIG:COUN 5
:OUTP ON
:READ?
Custom (list) sweep:
*RST
:SOUR:VOLT 0
:SOUR:DEL 0.1
:SOUR:SWE:RANG BEST
:SOUR:VOLT:MODE LIST
:SOUR:LIST:VOLT 1, 0, 1, 0, 1, 0
:TRIG:COUN 6
:OUTP ON
:READ?
*SourceMeter must be addressed to talk after sending :READ? and :CALC:DATA? to acquire data.
Comments
See Figure 6
Restore GPIB defaults (source V, measure I).
Set bias level to 0V.
Set delay to 100ms.
Select best source ranging.
Select the sweep source mode.
Select the linear sweep scale.
Set start level to 1V.
Set stop level to 10V.
Set step level to 1V.
Set trigger count to 10.
Turn output on.
Trigger sweep and acquire data.
See Figure 7
Restore GPIB defaults (source V, measure I).
Set bias level to 0V.
Set delay to 100ms.
Select best source ranging.
Select the sweep source mode.
Select the log sweep scale.
Set start level to 1V.
Set stop level to 10V.
Set number of sweep points to 5.
Set trigger count to 5.
Turn output on.
Trigger sweep and acquire data.
See Figure 8
Restore GPIB defaults (source V, measure I).
Set bias level to 0V.
Set delay to 100ms.
Select best source ranging.
Select the list source mode.
Specify source list (1V, 0V, 1V, 0V, 1V and 0V).
Set trigger count to 6.
Turn output on.
Trigger sweep and acquire data.